image dons

je fais un don

STENTYS Self-Apposing® sirolimus-eluting stent in ST-segment elevation myocardial infarction : results from the randomised APPOSITION IV trial

Bourses ACTION 2025 - Candidatures ouvertes !
M2, thèse de sciences, mobilité aux USA, soins infirmiers, prix de thèse de médecine - Date limite de candidatures : 06 avril 2025

Colloquium "La cardiologie préventive : nouveautés" (mardi 1er avril 2025)
Réservez vos places !

Publié dans EuroIntervention 2016 Feb;11(11):e1267-74

Auteurs : van Geuns RJ, Yetgin T, La Manna A, Tamburino C, Souteyrand G, Motreff P, Koch KT, Vrolix M, IJsselmuiden A, Amoroso G, Berland J, Montalescot G, Teiger E, Christiansen EH, Spaargaren R, Wijns W.

Article disponible en consultant le site

Abstract

EuroIntervention

Background and aim :

We sought to investigate the impact of the self-apposing, sirolimus-eluting STENTYS stent on midterm and long-term stent apposition and strut coverage compared with a zotarolimus-eluting balloon-expandable stent in ST-segment elevation myocardial infarction (STEMI) patients undergoing primary percutaneous coronary intervention (PPCI).

Methods :

In the APPOSITION IV trial, 152 STEMI patients were randomised (3:2) to the self-apposing, sirolimus-eluting STENTYS stent or a commercially available zotarolimus-eluting balloon-expandable stent at 12 sites in five countries with angiographic follow-up and optical coherence tomography at four or nine months.

Results :

At four months, a lower percentage of malapposed stent struts was observed in the STENTYS group (N=21; Nstruts=501) compared with controls (N=26; Nstruts=326; 0.07% vs. 1.16%; p=0.002) with significantly more covered struts, using a 20 µm cut-off (94.32% vs. 89.09%; p=0.003). At nine months, the primary endpoint (percentage malapposed stent struts) was similar in both groups (STENTYS, N=40; Nstruts=566; control, N=21; Nstruts=292), showing complete apposition (p=0.55) and near total (>96%) coverage (p=0.58).

Conclusion :

In STEMI patients undergoing PPCI, the self-apposing, sirolimus-eluting STENTYS stent was equivalent to a conventional drug-eluting balloon-expandable stent with respect to late stent strut apposition and coverage at nine months. However, stent strut apposition and coverage at four months were significantly better in the STENTYS group.

Autres actualités

+

01/10/2023


Rationale and design of the ARAMIS trial: Anakinra versus pl...

Arch Cardiovasc Dis. 2023 Oct;116(10):460-466
+

01/09/2023


Comparison of three echo-guidance techniques in percutaneous...

Arch Cardiovasc Dis. 2023 Sep 29:S1875-2136(23)00171-7